Executive Summary
An FDA inspection of computerized systems does not follow a separate computer checklist. Investigators work from the compliance program for the type of inspection, which asks about the validation and security of computerized or automated processes inside each of the six CGMP systems, and from the Investigations Operations Manual, which tells them how to handle electronic records once they are at your site. Both were checked against FDA’s own pages in October 2026. The 2026 IOM is current, the preapproval inspection program was revised in June 2026, and the main surveillance program for drug manufacturing still dates from 2022.
The rule that matters most for preparation is in the IOM. Outside of certain bioresearch monitoring inspections (clinical and nonclinical studies) and certain biologics inspections, investigators are told not to operate your system themselves. They watch your people operate it, ask them to run queries, and assume any custom query is unvalidated until shown otherwise. Your staff at the keyboard are therefore the interface between the investigator and your data, and FDA’s own data integrity guidance counts personnel as part of “computer or related systems.”
This article sets out what is in force, how computerized systems show up in each kind of inspection, and how to prepare the walkthrough itself: who demonstrates, how to run a live audit trail review, how to show user access and administrator rights, what backup evidence to have ready, and the traps that turn a routine demonstration into a Form FDA 483 observation.
What Is in Force Today
Before planning a walkthrough, it helps to be precise about which FDA documents govern an FDA inspection of computerized systems right now. There is a lot of commentary on this subject, and much of it describes documents that have since been revised or that never applied to drug manufacturing in the first place. The list below was checked against FDA’s own pages and the PDFs themselves.
The Documents Investigators Work From
FDA compliance programs are instructions to FDA staff. They tell investigators what to cover and how to report it. FDA publishes them so industry can see the approach, and FDA states on its index page that they do not bind FDA or the public.2 For drug manufacturing, the documents that matter most for computerized systems are these.
| Document | Current version | What it says about computerized systems |
|---|---|---|
| Compliance Program 7356.002, Drug Manufacturing Inspections | Issued September 16, 2022; implemented October 17, 2022 | Lists validation and security of computerized or automated processes as a coverage item in the facilities and equipment, materials, production, packaging and labeling, and laboratory systems1 |
| Compliance Program 7356.002F, Active Pharmaceutical Ingredient Process Inspection | Issued August 1, 2025; implemented September 2, 2025 | Names systems that allow alteration or deletion of raw data, and systems without audit trails activated, as examples of data integrity failures that may lead to an Official Action Indicated classification3 |
| Compliance Program 7346.832, Preapproval Inspections | Issued June 29, 2026; implemented August 10, 2026 | Keeps the data integrity audit as one of four inspection objectives, to be covered on every preapproval inspection4 |
| Investigations Operations Manual (IOM) | 2026 edition | Chapter 5 sets the rules for requesting, receiving, and handling electronic records during an inspection56 |
| Data Integrity and Compliance With Drug CGMP: Questions and Answers | December 2018, final | Explains how FDA reads 21 CFR 211.68 and related rules on access, shared logins, audit trail review, and backup7 |
| 21 CFR 211.68 | In force | The CGMP requirement for controls over computer or related systems, input and output checks, and backup files8 |
| Part 11 Scope and Application guidance | August 2003, final | FDA’s statement of enforcement discretion for certain Part 11 provisions, with enforcement through the underlying predicate rules9 |
Two points in that table are worth stating plainly. First, the main surveillance program for finished drug manufacturing, 7356.002, has not been reissued since 2022. FDA’s index page was updated in September 2026, and it still links the 2022 version.2 Second, the Part 11 enforcement discretion described in the 2003 guidance is still FDA’s stated position. FDA said in that guidance that it would not take enforcement action on the validation, audit trail, record retention, and record copying requirements of Part 11 as explained there, but that records must still meet the predicate rules and FDA can act on violations of those rules.9 That is why drug manufacturing warning letters about computer controls cite 21 CFR 211.68(b) and the laboratory record rules far more often than Part 11 itself.
What Changed in 2025 and 2026
There was no new FDA document in 2025 or 2026 that set out a new method for reviewing computerized systems in drug manufacturing. What changed is mostly about how and when investigators arrive, and how much they can see before they arrive.
- Unannounced foreign inspections. On May 6, 2025, FDA announced it would expand unannounced inspections at foreign manufacturing facilities, building on a pilot in India and China. The announcement said FDA conducts about 12,000 domestic and 3,000 foreign inspections each year.11 For a foreign site, the practical result is that the walkthrough team may have no notice at all.
- Remote regulatory assessments made final. FDA finalized its question and answer guidance on remote regulatory assessments in June 2025. It says FDA may review electronic systems and source records through screen sharing and livestream or video, and may ask for read-only access to electronic databases or ask a firm to walk FDA through information in its database.10 The same guidance states that a remote regulatory assessment is not an inspection under section 704(a)(1) of the FD&C Act.
- A revised preapproval program. The June 2026 revision of 7346.832 describes a stronger risk-based approach to deciding whether a preapproval inspection is needed. It allows a records request under section 704(a)(4) to be used in place of, or ahead of, a preapproval inspection.4
- A revised API program. The August 2025 API program added ICH Q9(R1), Q10, and Q12 elements and kept explicit data integrity examples, including systems without audit trails activated.3
- A 2026 IOM. FDA posted the 2026 edition of the Investigations Operations Manual, issued by the Office of Inspections and Investigations (OII), which is the name FDA’s field inspection organization now uses.5
Status note. The FDA guidance on Computer Software Assurance, issued in its current form on February 3, 2026, is device guidance from CDRH and CBER.18 Drug companies use its risk-based thinking by analogy, but none of the drug compliance programs or the IOM chapter discussed in this article cites it. When your team prepares for a CGMP inspection, the predicate rules, the compliance program, and the IOM are the documents to read first.
Where Computerized Systems Fit in the Inspection Plan
The most common planning mistake is to treat computerized systems as a separate topic that an investigator will “get to.” In practice, systems come up inside whatever process the investigator is reviewing. A batch record review becomes a manufacturing execution system review. A stability question becomes a chromatography data system question. An environmental monitoring question becomes a particle counter question.
The Six-System Model in Surveillance Inspections
Compliance Program 7356.002 organizes surveillance inspections around six systems: quality, facilities and equipment, materials, production, packaging and labeling, and laboratory controls. A full inspection covers at least four of them, one of which must be the quality system. An abbreviated inspection normally covers at least two, again including the quality system.1
Computerized systems appear across those six. The program lists “Validation and security of computerized or automated processes” under the laboratory control system. It lists process validation “including validation and security of computerized or automated processes” under production, qualification, validation, and security of computerized or automated processes under materials, and validation and security of computerized processes under packaging and labeling.1 The program’s examples of findings that can support regulatory action include a lack of validation of computerized processes in several of those systems.
The practical point: the system that gets demonstrated is the one attached to the record the investigator picked. You do not choose. That is why walkthrough readiness has to cover every system that touches a CGMP record, not just the three or four large platforms that IT thinks of first. Building that list is its own discipline, covered in our article on a defensible system inventory.
The Data Integrity Audit in Preapproval Inspections
Preapproval inspections are where electronic records get the most direct attention. The June 2026 program sets four objectives: readiness for commercial manufacturing, conformance to the application, the data integrity audit, and commitment to quality in pharmaceutical development. For the data integrity audit, the program says “Cover on every PAI.”4
The program tells investigators to audit and verify raw data at the facility against the data summaries filed in the CMC section of the application. It names finished product stability, dissolution, content uniformity, and API impurity data as good candidates. It also draws a distinction that matters for how you prepare: information with factual integrity corresponds directly to what was submitted, such as a chromatogram whose peak area calculates to the reported assay value, while information with contextual integrity supports the submitted data, such as a chromatographic sequence showing all the assayed samples and no hidden failing results.4
For a walkthrough, that distinction translates into two different demands. Factual integrity means your team must be able to go from a number in Module 3 to the raw data file that produced it, quickly and without a detour through someone’s personal folder. Contextual integrity means they must be able to show everything around that file: the full sequence, the other injections, the reprocessing history, and the audit trail entries that explain them.
Clinical Systems in Bioresearch Monitoring Inspections
Sponsors and contract research organizations are inspected under Compliance Program 7348.810. Its section on electronic records tells investigators to describe the electronic systems used to manage critical data, including whether each system was owned and controlled by the sponsor or a third party, and how data entered it. It repeats the Part 11 enforcement discretion position and refers investigators to the IOM for read-only access.13
FDA’s October 2024 question and answer guidance on electronic systems in clinical investigations lists what FDA will generally focus on during a sponsor inspection. The list includes processes to ensure that only authorized individuals get appropriate access, change control for the system once in use, and contracts with IT service providers.14 The same guidance says it is the regulated entity’s responsibility to make validation documentation available on request, including documentation created and maintained by the IT service provider. If your EDC or IRT vendor holds the evidence, your walkthrough plan has to include how you will get it on the day.
The IOM Rules Behind Every System Walkthrough
The compliance programs tell investigators what to look at. The IOM tells them how to handle what they find. Chapter 5 of the 2026 IOM has a section on electronic records that every walkthrough team should read, because it explains why investigators behave the way they do at the screen.6
Investigators Watch; Your People Operate
The IOM instruction is direct: “Do not personally access a firm’s system to review electronic records, databases, or source data during the course of an inspection,” with two exceptions. One is a bioresearch monitoring inspection that meets the IOM criteria for read-only access. The other is a biologics inspection where it is not practical to oversee firm personnel while they access the system.6
When an investigator needs to see data, the IOM tells them to oversee the firm’s personnel accessing the system and have them answer questions, to ask the firm to run queries specific to the information of interest, to ask for the parameters used to generate the data, and to ask the firm to send the data securely or provide it on electronic media.6 It also tells investigators not to use the firm’s equipment or personnel to do computerized data manipulation for real-time review and analysis.
This one instruction shapes the entire walkthrough. In a drug manufacturing inspection, the investigator will almost never touch your keyboard. Your analyst, operator, or system owner will. Everything they do is visible, much of it is recorded in the audit trail under their own user ID, and every answer they give is something the investigator can write down. FDA’s data integrity guidance defines “computer or related systems” to include hardware, software, networks, cloud infrastructure, and “personnel, and associated documents.”7 The person demonstrating is part of the system being inspected.
Your Query Is Presumed Unvalidated
The IOM tells investigators that a custom report query, such as all complaints from the last 12 months with specific fields, may be needed to answer a request. It then says: “You must assume the query logic is not validated and take appropriate action to ensure the data is accurate and no data has been omitted due to a programming logic error occurring at the firm.”6
Before requesting data, investigators are told to find out which program holds it, what fields the firm routinely captures, whether fields contain coded values that need a lookup table, and whether linked tables in a relational database should also be included. The IOM suggests they may ask for a printout of the data structure or watch staff enter data at a terminal.6
Read from the firm’s side, that means three things. Your staff should be able to explain where each field in an export comes from. You should have a data dictionary or code list for any system that stores coded values. And if you run a custom query during the inspection, you should record the exact parameters you used and be ready to show that the result is complete, because the investigator has been told not to take that for granted.
Copies, Originals, and Verification
The IOM distinguishes an original copy of an electronic record, which is preserved to support observations or evidence, from a working copy used for review. It tells investigators to verify that copies they receive are accurate representations of the source records so they can later testify to that, and to document when, where, and from whom copies were obtained.6 It also tells investigators to check, before closing the inspection, that the files they received contain what was requested and can be opened.
Investigators may also take photographs. The IOM says they should not ask permission to take photographs, because photography is part of FDA’s inspection authority under section 704(a)(1).6 A screen showing a shared login prompt, an operating system clock, or an instrument setting page can be photographed as easily as a cracked floor.
When a Slow Answer Becomes a Finding
FDA’s June 2024 guidance on delaying, denying, limiting, or refusing an inspection describes behaviors FDA may treat as limiting an inspection under section 501(j) of the FD&C Act, which can make a drug adulterated. Several of its examples are about electronic records. They include omitting or limiting data when providing electronic copies, such as removing data columns in Excel, exporting data into reports without all the data fields unless FDA asked for that, and “locking the electronic worksheet so that the data cannot be searched, sorted, or analyzed by FDA.”12 The guidance also lists failing to produce requested records within the timeframe FDA asked for, without a reasonable explanation, as a form of delay.
The 2026 IOM tells investigators that if management delays producing records without a reasonable explanation, or limits their ability to copy records, they may call attention to section 501(j).6 Reasonable explanations listed in the guidance include records needing translation, records stored off-site, and a large volume of records.12 “The person who knows how to run that report is on vacation” is not on the list.
What this means in practice. A protected PDF export, a spreadsheet with hidden columns, or a summary report in place of the native data set can each be read as limiting the inspection. If you are asked for electronic data, provide it in a form FDA can search, sort, and analyze, and say clearly what the file contains and how it was produced.
Who Demonstrates the System
Because the investigator watches rather than drives, the choice of who is at the keyboard is one of the most important decisions in walkthrough preparation. It is also one of the decisions most often made on the morning of the inspection.
The Person at the Keyboard Should Be a Routine User
The default choice at many sites is the system administrator or an IT lead, because they know the system best. That choice creates problems. An administrator account on a production GxP system is the account most likely to have rights the investigator will ask about. Every action taken under it during the walkthrough appears in the audit trail. And an administrator who demonstrates a routine task invites a reasonable question: why can this person, who is not a reviewer or analyst, see and change CGMP data?
The better choice is the person who does the work every day under their own ordinary account: the analyst who runs the chromatography sequences, the operator who executes the electronic batch record, the QA reviewer who approves records in the eQMS. They know the workflow as it is performed, not as it was designed. When the investigator asks “show me how you would do this,” the answer comes from someone who does it.
The administrator still has a role. They should be in the room or on call, ready to show configuration settings, role definitions, or audit trail settings that a routine user cannot see. When they do, it should be clear to everyone why the elevated account is being used, and the use should be documented.
Roles on the Walkthrough Team
| Role | What they do during the walkthrough | Common mistake |
|---|---|---|
| Demonstrator | Operates the system under their own routine account and answers workflow questions | Choosing whoever is free rather than someone who does the task daily |
| System owner | Explains intended use, validation status, configuration decisions, and open deviations or changes | Answering questions about workflow they have never performed |
| System administrator | Shows configuration, role definitions, audit trail settings, and backup records when asked | Demonstrating routine tasks under an administrator account |
| QA data reviewer | Shows how audit trails are reviewed as part of record review and where that review is documented | Not being present, so nobody can explain the review evidence |
| Scribe | Records every request, every query and its parameters, and every file provided | Writing summaries instead of the exact request and response |
| Back room lead | Retrieves documents, coordinates vendor support, and checks exports before they leave the room | Editing or “cleaning up” an export before it is provided |
One Story, Told the Same Way
A walkthrough often goes wrong when different people give different answers to the same question. The analyst says audit trails are reviewed with every run. The QA reviewer says they are reviewed monthly. The SOP says something else again. Each answer may be partly true, but the investigator now has a documented inconsistency.
The fix is not a script. Scripts sound rehearsed and fall apart under follow-up questions. The fix is to make sure the procedures match what people do, and then to walk the team through the procedures before the inspection so that everyone knows what the documented answer is. If practice differs from the procedure, the time to find that out is during a rehearsal, when you can open a deviation or a change, not when an investigator is writing it down.
Answering Questions Without Guessing
The demonstrator should answer what they know, show what they can, and say plainly when a question is outside their knowledge. “I don’t know, but our system owner can answer that” is a good answer. A confident guess about how backups work, offered by an analyst who has never seen the backup configuration, is a poor one. It may be wrong, and if it is, the correction later looks like a change of story.
Live Audit Trail Review at the Screen
The single most predictable request in a computerized system walkthrough is to see the audit trail for a specific record. Investigators rarely ask to see an audit trail in general. They pick a batch, a sample, or a result, often one they selected from a summary, and ask to see what happened to it.
What FDA Expects the Audit Trail to Show
FDA’s data integrity guidance defines an audit trail as a secure, computer-generated, time-stamped electronic record that allows reconstruction of the events relating to the creation, modification, or deletion of an electronic record. Its example for an HPLC run is specific: the audit trail should include the user name, date and time of the run, the integration parameters used, and details of any reprocessing, with a justification for the reprocessing documented.7
On who reviews audit trails, the guidance says “Audit trail review is similar to assessing cross-outs on paper when reviewing data,” and that personnel responsible for record review should review the audit trails that capture changes to the data as they review the rest of the record.7 On frequency, it says that where CGMP rules set a review frequency for the data, such as review before batch release, the same frequency applies to the audit trail. Where they do not, the firm should set a frequency using risk assessment.
That gives the investigator a simple test to run at the screen. Pick a released batch. Ask to see the audit trail for its key records. Ask who reviewed it, when, and where that review is documented. If the answer is that audit trails are reviewed separately on a quarterly schedule, the follow-up question is how the batch was released before its audit trail was reviewed.
Preparing the Demonstration
Know where each audit trail lives
Many systems keep more than one audit trail: a record-level trail, a system or configuration trail, and sometimes an operating system log. The demonstrator should know which one answers which question and how to open each.
Practice filtering by record, user, and date
An audit trail with ten thousand rows is not a demonstration. The demonstrator should be able to filter to the record the investigator chose, to a date range, or to one user, and show the filter settings on screen.
Be able to explain every event type
Reason codes, system-generated events, failed login attempts, and reprocessing entries all need a plain explanation. If the team cannot explain an entry in rehearsal, they will not explain it well in front of an investigator.
Show the review evidence next to the trail
Have the record that documents the audit trail review for the same batch or result ready to open. The investigator will want to connect the review to the data, not hear that review happens.
Export completely, or not at all
If the investigator asks for a copy, export the full trail for the scope requested, in a searchable format, and record the parameters used. A partial export with columns removed is the behavior the delay guidance describes.
Trial Injections and Elevated Roles
The audit trail is where unexplained activity shows up, and investigators know where to look. A September 2026 warning letter to an API manufacturer described thousands of unexplained injections in one chromatography data system over 30 months and more than 70 trial injections in a second system in a single month. It said injections in one system were run under a “Developer” role rather than the “Operator” role without appropriate documentation, and that the firm’s own response stated nearly 20 percent of trial injections in that system were acquired or altered in that mode.17
FDA asked that firm for an independent data integrity assessment covering all computerized systems used in CGMP operations at all its facilities, “not limited to Empower and TotalChrom.”17 That request shows how a single demonstration can expand. Investigators who find a problem in one system are likely to ask whether the same control gap exists in others.
For the walkthrough, the preparation step is to look at your own audit trails before an investigator does. Search for injections that do not link to a sample, for activity under elevated roles during routine work, and for reprocessing without a documented reason. If you find any, investigate them through your normal quality system. An open, well-documented investigation is a far better position than a discovery made at the screen.
Do not tidy up the audit trail. Audit trails are meant to be permanent. Any attempt to remove, suppress, or reconfigure audit trail settings before an inspection is itself a serious data integrity problem, and the system or configuration audit trail will usually record it. If you find something you do not like, investigate it and document the outcome.
User Access and Administrator Rights
Access control is the area where a walkthrough most often produces an observation, because the evidence is right there on the screen. The question is simple: can the investigator see that only authorized people can change CGMP records, and that each action can be traced to one person?
The Rule and FDA’s Reading of It
21 CFR 211.68(b) says: “Appropriate controls shall be exercised over computer or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel.”8
FDA’s data integrity guidance explains how FDA reads that sentence. It recommends restricting the ability to alter specifications, process parameters, data, or methods by technical means where possible. It says “The system administrator role, including any rights to alter files and settings, should be assigned to personnel independent from those responsible for the record content.” And it says manufacturers should document authorized personnel’s access privileges for each CGMP computer system, for example by keeping a list of authorized individuals.7
On shared logins, the guidance says that when credentials are shared, a unique individual cannot be identified through the login, and the system would not conform to CGMP. It allows shared read-only accounts for viewing data, but says they do not meet the requirement for actions such as second-person review to be attributable to a specific individual.7
What the Warning Letters Show
Recent letters show what investigators find when they look at access at the screen.
- A March 2025 warning letter to a finished drug manufacturer cited 211.68(b) after investigators documented numerous deleted electronic raw data files and found that analysts had administrative rights capable of altering and deleting data, files, and folders on chromatographic systems, including the date and time of tests.15
- The September 2026 warning letter to an API manufacturer discussed earlier also described a laboratory analyst using another analyst’s login credentials to perform calibrations and sample testing, on a system that enforced individual, unique usernames and passwords. The firm’s own procedure already prohibited credential sharing. FDA found the firm’s retrospective review too narrow because it covered only 12 stand-alone analytical instruments over about one year and did not cover all computerized equipment that requires a login.17
Neither finding required a sophisticated review. One came from looking at what an ordinary user could do on an instrument computer. The other came from comparing who was logged in with who did the work. That is why the walkthrough has to cover standalone equipment and instrument PCs, not just enterprise applications, and why a procedure that forbids shared logins is not evidence that nobody shares them.
The Access Evidence to Have Ready
Current list of authorized users
For each system, a list of named users and their roles, matched to the people who work there today. Former employees and contractors should show as disabled, with dates.
Role and privilege definitions
A document or configuration report showing what each role can do, especially who can delete, reprocess, change methods, change audit trail settings, or change the system clock.
Who holds administrator rights
The names of administrators for the application and for the operating system of each instrument PC, and evidence that they are independent from the people who generate or approve the data.
Periodic access review records
The most recent review for each system, what it found, and what was changed as a result. An access review that never removes anyone is a weak piece of evidence.
Service, vendor, and generic accounts
Every account that is not tied to one named person, why it exists, who controls it, and how its use is monitored. These are the accounts investigators ask about.
Instrument computer settings
For instruments with local data storage: whether users log in individually, whether they can change the clock, and whether they can delete files from local folders.
A useful rehearsal is to have someone outside the system team sit down at each instrument computer with an ordinary user account and try the things an investigator would ask about: change the date, open the data folder, delete a test file, change an alarm limit. If any of those works, you have found your observation before the investigator does.
Backup and Restore Evidence
Backup is the part of 211.68(b) that walkthrough teams most often treat as somebody else’s problem. IT runs the backups, the vendor hosts the system, and nobody in the room can show that the data could be recovered. Investigators ask anyway.
What the Regulation and Guidance Require
21 CFR 211.68(b) requires that a backup file of data entered into the computer or related system be maintained, and that backup systems be designed to assure that backup data are exact and complete and secure from alteration, inadvertent erasures, or loss.8
FDA’s data integrity guidance says it uses “backup” in 211.68(b) to mean a true copy of the original record kept securely throughout the record retention period. The backup should contain the data, including associated metadata, in the original format or a compatible one. The guidance adds: “Temporary backup copies (e.g., in case of a computer crash or other interruption) would not satisfy the requirement in § 211.68(b) to maintain a backup file of data.”7
For clinical systems, FDA’s October 2024 guidance says that when records exist only in electronic form, backup and recovery procedures should protect against data loss, records should be backed up regularly, and backups should be stored in a secure location separate from the original records.14
How Backup Gaps Show Up
An August 2026 warning letter to a drug manufacturer making sterile products described integrity testing equipment whose reports were not saved with unique file names, so the data was overwritten after a set number of tests. FDA redacted that number in the published letter. The letter says the firm therefore had no means of tracking or retrieving records of failed leak tests, and it placed the finding under the citation for incomplete laboratory records.16
That example is useful because it involved neither a server failure nor a cyberattack. It involved the way one instrument saved its own reports. Many sites have equipment like that, and many have never checked what an instrument does to the previous file when it saves a new one, or what a vendor engineer does to local data during a service visit. A backup schedule on the server does not help if the original was overwritten on the instrument before the backup ran.
The Backup Evidence to Have Ready
| Evidence | What it should show | Where it often fails |
|---|---|---|
| Backup procedure or configuration | What is backed up, how often, where it goes, and how long it is kept | Covers servers but not instrument PCs that store data locally |
| Backup success records | Job logs or reports for a recent period, with failures followed up | Failures logged by IT but never reported to the system owner |
| Restore test record | A documented restore of real GxP data, with a check that the restored record and metadata match | Restore tested once at validation and never again |
| Hosted system evidence | Supplier backup and restore commitments in the quality agreement, plus evidence the supplier tests them | No way to get the evidence from the supplier during the inspection |
| Vendor service controls | A requirement that data be secured before service and checked after | Service visits handled as maintenance only, with no data step |
The demonstrator does not need to run a restore during the inspection. They need to be able to open the most recent restore test record for the system being discussed and explain it. If the system is hosted, the system owner needs to know where the supplier’s evidence is and how quickly it can be produced.
Common Traps and How to Rehearse for Them
Most problems in a computerized system walkthrough are not caused by bad systems. They are caused by ordinary decisions made in a hurry. The traps below come up again and again, and each one follows from the IOM rules and FDA guidance discussed above.
Traps at the Screen
Demonstrating in the wrong environment. Showing a training or validation environment without saying so. If the investigator later finds the production system behaves differently, the demonstration looks misleading. Always say which environment is on screen.
Logging in as an administrator. Using an elevated account for a routine task because it is faster. It raises the question of why that account can touch CGMP data, and the audit trail records it.
Exporting less than the full record. Providing a report without all the data fields, a locked file that cannot be searched or sorted, or a spreadsheet with columns removed. The 2024 guidance lists these as ways of limiting an inspection.
Running an ad hoc query without recording it. The investigator has been told to assume the query logic is not validated. If nobody wrote down the parameters, nobody can show the result is complete.
Saying a report cannot be produced. If a system can produce the data, saying it cannot is a credibility problem. If it cannot, explain why and offer another way to get the same information.
Guessing. A confident wrong answer from a demonstrator about backups, access, or configuration becomes an inconsistency the moment someone else gives the right one.
Traps Before the Investigator Arrives
- Assuming you will have notice. FDA’s May 2025 announcement made unannounced inspections at foreign sites more likely, and FDA said in the same announcement that domestic inspections are pre-announced only in specific programs and cases.11 Readiness that depends on a two-week warning is not readiness.
- Assuming the first look will be on site. A records request under section 704(a)(4) can come before an inspection or, in a preapproval case, in place of one.4 The electronic records you send in response are evaluated with the same care as records shown on site.
- Assuming the vendor will be available. For hosted systems, the supplier holds part of the evidence. FDA’s clinical guidance puts responsibility for that evidence on the regulated entity.14 Agree in the quality agreement how fast the supplier will support an inspection, and test it.
- Assuming procedure and practice match. The most common source of inconsistent answers is a procedure written for the system as designed, followed by staff working the system as it is.
- Forgetting standalone equipment. The 2025 and 2026 letters cited above all involved instrument-level controls: chromatographic systems where analysts had administrative rights, stand-alone analytical instruments where one analyst used another’s credentials, and test equipment that overwrote its own reports.151716
A Rehearsal That Tests the Right Things
A good rehearsal does not walk through a prepared demonstration. It reproduces the way an investigator works. Someone playing the investigator picks a record from a summary list, such as a released batch or a stability result, and asks the team to show where it came from. They then ask the follow-up questions the IOM and FDA guidance suggest.
Pick the record, not the system
Start from a CGMP record and let the system follow. This tests whether the team can find the right system, the right data, and the right person without notice.
Trace it to raw data
For a laboratory result, open the raw data file, the full sequence, and the processing history. For a batch, open the electronic batch record and its exceptions.
Open the audit trail and the review
Filter the audit trail to the record, explain each event, and open the record of the audit trail review that preceded release or approval.
Ask who else could have changed it
Show the list of users with rights to change that record, the administrators, and the last access review. Check whether an ordinary user can change the clock or delete files.
Ask how it would be recovered
Open the backup configuration for the system and the most recent restore test. For hosted systems, time how long it takes to get the supplier’s evidence.
Request a full export
Ask for the data set behind a summary, in native or searchable format. Check that it is complete, unlocked, and documented with the parameters used.
Record what went wrong in the rehearsal as you would record any other finding. If the team could not find the restore record, that is a gap in documentation or ownership. If two people gave different answers about audit trail review frequency, that is a gap between procedure and practice. Treat each one through the quality system, with an owner and a date.
A sign the rehearsal worked. The best outcome is not a smooth demonstration. It is a short list of specific gaps, each with an owner, found by your own people at a time of your choosing. A rehearsal that finds nothing usually means the person playing the investigator was too polite.
Questions Leaders Should Ask Their Teams
Senior leaders rarely sit at the keyboard during an inspection, but they decide whether the preparation happens. A few direct questions will show where a site stands:
- For our three most inspected record types, who would demonstrate the system, and do they use it every day under their own account?
- Can we show the audit trail review for the last released batch, and does it match what our procedure says?
- Which of our instrument computers let an ordinary user change the date or delete data, and how do we know?
- When was the last time we restored real GxP data from backup for each critical system, and where is the record?
- For our hosted systems, how long would it take to get backup, access, and audit trail evidence from the supplier during an inspection?
- If we received a 704(a)(4) records request for electronic data tomorrow, who would prepare it, and how would we check it is complete?
Conclusion
FDA’s approach to computerized systems in drug inspections has not been rewritten in 2025 or 2026. The surveillance program still asks about validation and security of computerized processes inside each CGMP system, the preapproval program still makes the data integrity audit an objective on every inspection, and the 2018 data integrity guidance still explains how FDA reads 211.68(b). What has changed is the setting: more unannounced foreign inspections, a final framework for remote assessments that includes screen sharing and database walkthroughs, and a revised preapproval program that can use records requests in place of a visit. The constant is the IOM rule that investigators watch while your people operate. That rule makes the walkthrough a test of your staff, your procedures, and your evidence at the same time.
Sakara Digital works with pharma and biotech organizations preparing their computerized systems, their records, and their people for this kind of scrutiny, from audit trail review practice to access and backup evidence for hosted and standalone systems. If you are planning an inspection readiness effort and want an independent perspective on where your walkthrough would hold up and where it would not, we are happy to have that conversation.
For Further Reading
For Further Reading
- Computerized System Inventory: Building One You Can Defend
- Audit Trail Review as a Data Quality Control: Two Pilots Every Pharma Site Should Run
- Access Reviews That Actually Catch Problems in GxP Systems
- Backup and Restore Testing for GxP Systems: The Drill Many Sites Skip
- Remote and Hybrid GxP Inspections in 2026: What Changed
References & Sources
- U.S. Food and Drug Administration. “Compliance Program 7356.002: Drug Manufacturing Inspections.” Date of issuance September 16, 2022; implementation date October 17, 2022. https://www.fda.gov/media/75167/download
- U.S. Food and Drug Administration. “Drug Compliance Programs.” Index page, content current as of September 16, 2026. https://www.fda.gov/drugs/guidance-compliance-regulatory-information/drug-compliance-programs
- U.S. Food and Drug Administration. “Compliance Program 7356.002F: Active Pharmaceutical Ingredient Process Inspection.” Date of issuance August 1, 2025; implementation date September 2, 2025. https://www.fda.gov/media/187860/download
- U.S. Food and Drug Administration. “Compliance Program 7346.832: Preapproval Inspections.” Date of issuance June 29, 2026; implementation date August 10, 2026. https://www.fda.gov/media/193382/download
- U.S. Food and Drug Administration, Office of Inspections and Investigations. “Investigations Operations Manual 2026.” Content current as of July 15, 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/inspection-references/investigations-operations-manual
- U.S. Food and Drug Administration. “Investigations Operations Manual 2026, Chapter 5: Inspections” (sections 5.5.2.3, 5.6.6, 5.6.7, 5.6.10, 5.13.2.5.1, and 5.14.2.1). https://www.fda.gov/media/166533/download
- U.S. Food and Drug Administration. “Data Integrity and Compliance With Drug CGMP: Questions and Answers.” Guidance for industry, December 2018. https://www.fda.gov/media/119267/download
- Electronic Code of Federal Regulations. “21 CFR 211.68: Automatic, mechanical, and electronic equipment.” https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-211/subpart-D/section-211.68
- U.S. Food and Drug Administration. “Part 11, Electronic Records; Electronic Signatures: Scope and Application.” Guidance for industry, August 2003. https://www.fda.gov/media/75414/download
- U.S. Food and Drug Administration. “Conducting Remote Regulatory Assessments: Questions and Answers.” Guidance for industry, June 2025. https://www.fda.gov/media/160173/download
- U.S. Food and Drug Administration. “FDA Announces Expanded Use of Unannounced Inspections at Foreign Manufacturing Facilities.” Press announcement, May 6, 2025. https://www.fda.gov/news-events/press-announcements/fda-announces-expanded-use-unannounced-inspections-foreign-manufacturing-facilities
- U.S. Food and Drug Administration. “Circumstances that Constitute Delaying, Denying, Limiting, or Refusing a Drug or Device Inspection.” Guidance for industry, June 2024, Revision 1. https://www.fda.gov/media/86328/download
- U.S. Food and Drug Administration. “Compliance Program 7348.810: Sponsors and Contract Research Organizations.” Implementation date September 15, 2021. https://www.fda.gov/media/75916/download
- U.S. Food and Drug Administration. “Electronic Systems, Electronic Records, and Electronic Signatures in Clinical Investigations: Questions and Answers.” Guidance for industry, October 2024. https://www.fda.gov/media/166215/download
- U.S. Food and Drug Administration. Warning Letter 320-25-52 to International Laboratories Corp, March 10, 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/international-laboratories-corp-698522-03102025
- U.S. Food and Drug Administration. Warning Letter 320-26-117 to Reliance Life Sciences Private Limited, August 18, 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/reliance-life-sciences-private-limited-730801-08182026
- U.S. Food and Drug Administration. Warning Letter 320-26-127 to Curia New York, Inc., September 18, 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/curia-new-york-inc-728790-09182026
- U.S. Food and Drug Administration. “Computer Software Assurance for Production and Quality Management System Software.” Guidance for industry and FDA staff, February 3, 2026. https://www.fda.gov/media/188844/download