What Print-and-Sign Looks Like on a Working Site

Ask a quality team whether they run hybrid systems and many will say no. They have a chromatography data system (CDS), a laboratory information management system (LIMS), an electronic quality management system (eQMS), and perhaps an electronic batch record. Then walk the lab and the quality offices and look at what comes off the printers. That is where the print-and-sign hybrid records are.

PIC/S defines a hybrid system as one that typically consists of an electronic system generating electronic data, supplemented by a defined manual system that generates a paper-based record, so that the complete data set consists of both electronic and paper data together.1 WHO uses a shorter definition: a combination of electronic systems and paper systems.2 Print-and-sign is the most common way a site creates one without meaning to.

Common Forms of the Habit

The pattern repeats across functions. Some of the forms we see most often:

  • Laboratory results. An analyst runs a sequence in the CDS, prints the summary report and chromatograms, signs the printout as performer, and passes it to a reviewer who signs it in ink. The electronic data stays on the server, but the review was done on paper.
  • Certificates of analysis and release documents. The LIMS generates a certificate, which is printed so the quality approver can sign it before it goes into the batch file.
  • Instrument printouts from networked equipment. A particle counter, a water system monitor, or a stability chamber controller stores electronic data, but the site files the paper strip or a printed trend as the official record.
  • Quality records. A deviation or change control is drafted in the eQMS, then printed for a wet signature from a department head who does not have an account or has never been trained on the approval workflow.
  • Spreadsheets. A calculation is done in a spreadsheet, the result is printed, and the printout is signed and attached to a worksheet. The spreadsheet file itself may or may not be kept.
  • Batch record exceptions. The electronic batch record handles most steps, but certain steps, such as data from a standalone instrument or a supervisor sign-off, are printed and signed outside the system.
  • Vendor and contractor reports. A contract lab sends a PDF report, which is printed, signed as “reviewed,” and filed, while the underlying electronic data stays at the contractor.

In each case a person has made a choice, often an unspoken one, that the paper is the record and the electronic data is a by-product. That choice is the root of most of the problems described later in this article.

When Paper Is the Right Original

Not every printout is a problem. The guidance draws a clear line between simple equipment and systems that store data. PIC/S notes that very simple electronic systems, such as balances, pH meters, or simple processing equipment that do not store data, generate directly printed paper records. For these, the original record should be signed and dated by the person generating it, with information such as sample ID and batch number recorded on it, and attached to the batch processing or testing record.1 FDA takes the same view: during data acquisition, pH meters and balances may create a paper printout as the original record, and that printout (or a true copy) must be retained.3

The line moves as soon as the equipment stores or processes data. FDA’s data integrity guidance uses the example of an FT-IR spectrometer: the spectral file is dynamic and can be reprocessed, and a static record or printout “is fixed and would not satisfy CGMP requirements to retain original records or true copies.”3 MHRA makes the same point in its definition of raw data: where data were captured in a dynamic state and generated electronically, paper copies cannot be treated as the raw data.4

A simple test for any printout on your site. Ask one question: does the system that produced this paper store, process, or allow reprocessing of the data? If the answer is no (a basic balance or pH meter with no memory), the signed printout can be the original record. If the answer is yes, the electronic data is the original record, and the printout is at best a copy. Signing the copy does not change which one is the original.

Why the Habit Persists

Print-and-sign is rarely a deliberate strategy. It is the residue of decisions made one system at a time, often years apart. Understanding why it persists matters, because each cause needs a different fix, and a plan that addresses only the technology will fail if the reasons are procedural or cultural.

The System Has No Electronic Signature Turned On

This is the most common technical reason, and it is often not a missing feature. Many commercial CDS, LIMS, and eQMS products include electronic signature functions, but the function was never licensed, configured, or validated at implementation. The original project scope covered data capture, and signing was left for a later phase that never started. Other systems, particularly older standalone instrument software, have no signature function at all. Telling these two situations apart is one of the first jobs in the inventory step below, because the first is a configuration and validation project and the second is a replacement decision.

Procedures Were Written for Paper

Even when a system can sign electronically, the standard operating procedures (SOPs) often still say “print the report, sign and date, attach to the worksheet.” Batch record templates still have a signature box for a step the system already records. Training materials still show the paper flow. Staff follow the procedure in front of them, so the habit continues long after the technical reason for it has gone. Any elimination plan has to include the document changes, not only the system changes.

Distrust of Electronic Records

Many experienced reviewers and approvers trust ink more than a login. Some worry that an electronic signature could be applied by someone who knows their password. Some feel that a reviewer has not done a proper review unless they have physically handled the pages. Some have lived through a system outage and want a paper copy “just in case.” These concerns are worth hearing, because some of them point to real control gaps, such as shared accounts or weak password rules. But the answer to a weak electronic control is a stronger electronic control, not a paper copy that has its own, larger weaknesses.

Resistance to change is not unique to signatures. In polling at the Pistoia Alliance’s April 2025 conference in London, resistance to change emerged as the biggest barrier to innovation, at 51 percent.5 A shift from ink to electronic signatures is a small change by technology standards, but it changes what reviewers do every day, which is why it meets the same resistance.

A Misreading of the 2003 Part 11 Guidance

FDA’s 2003 guidance on the scope and application of Part 11 is sometimes read as permission to print and sign. The guidance does say that when people use computers to generate paper printouts of electronic records, and those paper records meet all predicate rule requirements and people rely on the paper to perform regulated activities, FDA would generally not consider them to be using electronic records in place of paper.6 It also says that paper and electronic record and signature components “can co-exist” in a hybrid situation, as long as predicate rule requirements are met and the content and meaning of the records are preserved.6

Two parts of the same guidance are often skipped. First, FDA says actual business practice decides whether you are relying on the electronic record, and that “the Agency may take your business practices into account in determining whether part 11 applies.”6 If the reviewer opens the CDS to check integration, or the electronic data is used for trending or investigations, the site is relying on the electronic record whatever the SOP says about paper. Second, the 2018 data integrity guidance later made clear that a printout of a dynamic record does not satisfy the requirement to keep original records or true copies.3 The 2003 guidance narrowed how FDA would enforce Part 11. It did not make the printout the complete record.

The Data Integrity Problems Print-and-Sign Creates

The problems with hybrid records are not theoretical. They show up in FDA warning letters, in MHRA and PIC/S inspection findings, and in internal audits. Most of them come back to one question that a hybrid process cannot answer cleanly: which version is the record?

1997Year Part 11 electronic signatures became usable, with certification covering signatures used on or after August 20, 19977
51%Resistance to change, the top innovation barrier in polling at a 2025 Pistoia Alliance conference5
15+Plastic waste bags of torn CGMP records, including balance printouts, described in one 2025 FDA warning letter8

Which Version Is the Record?

In a hybrid process there are two candidates for “the record”: the electronic data in the system and the signed paper in the file. They are supposed to match, but nothing forces them to. An analyst can reprocess a chromatogram after printing it. A sequence can contain injections that never appear on the printed summary. A spreadsheet can be edited after its printout was signed. When the two diverge, the site has to explain which one it relied on, and why the other one says something different.

FDA’s July 2026 warning letter to Shimoga Chemicals, an API manufacturer, shows what happens when the paper and the electronic data tell different stories. FDA found unreported HPLC injections that were not documented in laboratory batch records, including one batch where an unreported injection gave an out-of-specification assay result while a different, passing result was documented in the batch record. Because the firm had no procedures for electronic data review and its quality unit did not review the electronic data, the batch was released to the U.S. market. In its response, the firm acknowledged undocumented trial injections, unreported data, discarded printouts, and inadequate review of electronic laboratory records.9 The paper record was complete on its face. The electronic record held the failing result.

Signatures That Are Not Linked to the Data

Part 11 does not only govern electronic signatures. Section 11.70 also covers handwritten signatures applied to electronic records, and requires that they be “linked to their respective electronic records to ensure that the signatures cannot be excised, copied, or otherwise transferred to falsify an electronic record by ordinary means.”10 A signature on a printout links to the printout. Unless the printout identifies exactly which electronic files, versions, and processing results it represents, there is no reliable link to the data the signer is said to have approved. R.D. McDowall, writing in LCGC, makes the point that many people assume Part 11 applies only to electronic signatures and overlook the requirement to link handwritten signatures to the underlying records.11

Audit Trail Gaps and Paper-Only Review

A reviewer who signs a printout has reviewed the printout. The audit trail, the processing history, the list of all injections in the sequence, and any changes to integration parameters are in the system, not on the paper. That is why MHRA says increased data review “is likely to be required for hybrid systems because they are vulnerable to non-attributable data changes.”4 PIC/S makes the same point and adds that procedures for reviewing hybrid data should explain how electronic data and paper-based data are correlated to form a complete record.1

McDowall traces this risk back to the Able Laboratories case, where falsified data went undetected because inspection attention was on paper printouts rather than the electronic records behind them.11 The lesson is not that reviewers are careless. It is that a paper-based review cannot see what the paper does not show.

Lost, Discarded, and Uncontrolled Paper

Paper is easy to lose and easy to throw away. FDA’s February 2025 warning letter to Granules India described a large amount of torn CGMP records found in at least 15 plastic waste bags during the inspection, including analytical balance printouts and worksheets containing manufacturing and testing data. Among the root causes the firm identified were unclear good documentation practice and data integrity protocols, and inadequate SOP provisions for attaching defective or illegible printouts with annotations.8 When a printout is the record, every reprint, misprint, and draft is a potential discarded original.

The April 2026 warning letter to Fareva Amboise shows the reverse problem: paper records that were never produced. FDA found that particle counter data showed multiple Grade A action-limit exceedances, but the firm could not produce original printed documentation of those failing results despite extensive searches, and only conforming retest results were included in production documentation. Management confirmed “that no controls were in place to ensure the integrity of printed data records.”12 The electronic data survived on the instrument. The paper record, which the site treated as official, did not include the failures.

Two Records to Keep for the Full Retention Period

Hybrid records also double the retention burden. The site must keep the electronic data readable for the full retention period and keep the paper safe, and it must keep the link between them. MHRA states that where hybrid records are stored, “references between physical and electronic records must be maintained such that full verification of events is possible throughout the retention period.”4 That is a demanding requirement when the paper is in an offsite archive and the electronic data has been migrated twice to new software versions. Many sites discover the broken links only when an investigation or inspection asks for a record from ten years ago.

What FDA, MHRA, PIC/S, and WHO Expect From Hybrid Systems

The major guidance documents agree more than they differ. None of them bans hybrid systems outright. All of them expect a site to define the complete record, control the link between paper and electronic parts, and review both. PIC/S and WHO go further and set a clear direction: replace hybrid systems.

SourceWhat it says about hybrid recordsWhat it means in practice
FDA, Data Integrity and Compliance With Drug CGMP Q&A (December 2018)3A printout of a dynamic record does not satisfy the requirement to retain original records or true copies. Electronic signatures with appropriate controls can replace handwritten signatures in any CGMP required record.The electronic data must be kept and reviewed. A handwritten signature is not required anywhere in the drug CGMPs if a controlled electronic signature is used.
FDA, Part 11 Scope and Application guidance (August 2003)6Paper and electronic components can co-exist if predicate rules are met. FDA recommends deciding in advance, and documenting, whether you rely on the paper or the electronic record.Write down, record by record, which format you rely on. Business practice, not the SOP alone, decides.
21 CFR 11.7010Handwritten signatures executed to electronic records must be linked to those records so they cannot be excised, copied, or transferred.A signed printout needs an unambiguous link to the exact electronic files it represents.
MHRA GXP Data Integrity Guidance (March 2018)4Document what constitutes the whole data set; review and retain all records in it. Expect increased data review. Maintain references between physical and electronic records for the full retention period.Each hybrid process needs a written definition of its complete record and a review that covers the electronic part.
PIC/S PI 041-1, Section 9.10 (July 2021)1Hybrid systems need specific and additional controls; their use is discouraged and they should be replaced whenever possible. Requires a system description and procedures for the paper-electronic interface and for review.Replacement is the stated direction. Until then, a documented system description and a hybrid review procedure are expected.
WHO TRS 1033, Annex 4 (2021)2Legacy hybrid systems should be discouraged with a priority timeline for replacement. Where hybrids are used, including for an interim period, the risk should be assessed and mitigating controls documented.A replacement timeline and a CAPA plan are expected, not open-ended use.
EU GMP Annex 11 (2011)13Electronic records may be signed electronically. Electronic signatures should be permanently linked to their record and include date and time. Batch certification in a computerized system should use an electronic signature.For EU-facing sites, electronic signing is the expected route for batch release in a computerized system.

The Direction of Travel Is Clear

PIC/S states that “the use of hybrid systems is discouraged and such systems should be replaced whenever possible.”1 WHO says “the use of legacy hybrid systems should be discouraged and a priority timeline for replacement should be established,” and, in its discussion of contemporaneous recording, that “the replacement of hybrid systems should be a priority with a documented CAPA plan.”2 CAPA here means corrective and preventive action, the formal quality process for fixing a problem and keeping it from coming back. Put together, these statements mean an inspector who finds a hybrid process will ask two questions: what controls are in place today, and when will it be replaced?

The EU is also revising Chapter 4 of its GMP Guide, and the July 2025 consultation draft contains new text on hybrid records and on signing electronic records electronically. That revision is still a draft, and we covered what it proposes in a separate article on the revised Chapter 4. This article does not depend on it. Everything in the plan below is already expected under guidance in force today.

Paper Is Not Banned, but It Must Be Defined

None of these documents says a site may never use paper. They say the site must know what its complete record is, keep it, review it, and keep the parts linked. A site that cannot answer “what is the complete record for this test?” in one sentence has a hybrid problem regardless of how much paper it uses. A site that can answer the question, and that has a dated plan to move the remaining paper into the system, is in a defensible position.

Step One: Inventory Every Hybrid Process

You cannot end a habit you have not mapped. The first step is a complete inventory of every place where a computerized system creates data and a person prints it to sign, review, or file. Most sites underestimate this list. The inventory is also the evidence an inspector will want to see, because it shows the site has defined and identified its hybrid systems, which is the first item PIC/S lists for inspectors to check in its section on hybrid systems.1

Where to Look

Use several sources, because no single one will find everything:

  • SOPs and work instructions. Search controlled documents for words such as “print,” “printout,” “attach,” “sign and date the report,” and “wet signature.” Each hit is a candidate.
  • Batch record and worksheet templates. Look for attachment fields and signature boxes for steps a system already performs.
  • The computerized system inventory. PIC/S expects regulated users to have an inventory of all computerized systems in use, with their function, criticality, and validation status.1 For each system, ask whether its outputs are printed for signature.
  • Printers. Walk the lab, the quality offices, and the production floor. Ask what each printer is used for. Instrument and lab printers are often the fastest way to find undocumented hybrid steps.
  • Archive intake. Ask the archivist which record types arrive as paper and which systems they came from.
  • Recent audits and deviations. Data integrity observations, missing printouts, and transcription errors often point to a hybrid step.

What to Record for Each Process

A short, consistent record for each hybrid process makes the later steps faster. The fields below are the minimum we recommend.

FieldWhat to captureWhy it matters
Process and record typeFor example, “HPLC assay release testing, CDS report and review”Defines the scope of the fix
Source system and versionThe system that creates the electronic dataDetermines whether electronic signing is possible
Electronic signature capabilityAvailable and in use; available but not configured; not availableSeparates configuration projects from replacement decisions
Current “official” recordWhat the SOP says is the record, and what people rely on in practiceExposes the gap the 2003 Part 11 guidance warns about
Static or dynamic dataCan the data be reprocessed or changed in the system?Dynamic data means the printout cannot be the original
GMP impact and data criticalityDirect product quality impact, indirect, or noneSets the priority order
Who prints and who signsRoles, and whether signers have system accountsShows the access and training work needed
Link between paper and dataDoes the printout name the file, sequence, version, and location?Tests compliance with 21 CFR 11.70
VolumeApproximate printouts per monthHelps plan effort and measure progress later

Sort the Inventory Into Four Groups

Once the inventory is complete, each process falls into one of four groups. The group decides the path.

Group A

Simple Equipment, Paper Is the Original

Balances and pH meters that store no data. Keep the signed printout as the original, attached to the batch or test record, as PIC/S and FDA describe. Check that thermal paper is copied before it fades. No elimination needed, but connecting the instrument to a system is worth considering when equipment is replaced.

Group B

E-Signature Available, Not Configured

The system can sign electronically, but the function was never turned on or validated. These are the quickest wins: a configuration change, a risk-based validation effort, SOP updates, and training. Prioritize by data criticality.

Group C

No E-Signature, Replacement Planned

Older standalone software with no signature function and no upgrade path. Put interim controls in place now and set a dated replacement plan, as WHO expects. Do not let “replacement planned” become permanent.

Group D

Habit Only

The system already signs electronically, but people still print and sign “for the file.” These need no system work at all. Change the SOP, stop the printing, and decide what to do with existing paper. They are often a large share of the list.

Decide the record of reliance now, even before anything changes. FDA’s 2003 guidance recommends that “for each record required to be maintained under predicate rules, you determine in advance whether you plan to rely on the electronic record or paper record to perform regulated activities,” and that you document the decision, for example in an SOP.6 Adding a “record of reliance” column to the inventory, and then writing those decisions into procedures, closes a gap many sites have today, even if every process stays hybrid for another year.

Step Two: Enable Part 11 Electronic Signatures

For Group B processes, the fix is to turn on and validate the electronic signature function in the system you already own. FDA has been explicit that this is acceptable throughout the drug CGMPs: “electronic signatures with the appropriate controls can be used instead of handwritten signatures or initials in any CGMP required record,” including master production and control records, where 21 CFR 211.186(a) refers to a full handwritten signature.3 The work is mostly about confirming the controls Part 11 requires and writing them down.

The Part 11 Controls to Confirm

Part 11 has a short list of signature requirements. Before going live, confirm and document each one for the system in question:

  • Signature manifestation (11.50). Signed records must show the printed name of the signer, the date and time of signing, and the meaning of the signature, such as review, approval, responsibility, or authorship. These items must be part of any human-readable form of the record, including a display or printout.14
  • Signature and record linking (11.70). The signature must be linked to its record so it cannot be removed, copied, or moved to falsify another record.10 PIC/S adds a practical test: if a signed record is changed later, the record should show the amendment and appear as unsigned until it is reviewed and signed again.1
  • Uniqueness and identity (11.100(a) and (b)). Each electronic signature must be unique to one person and never reused or reassigned. The organization must verify the person’s identity before it assigns or sanctions an electronic signature.7
  • Signature components (11.200). Signatures not based on biometrics must use at least two distinct components, such as a user ID and password. The first signing in a continuous session uses all components; later signings in the same session use at least one component that only the owner can execute.7
  • Password and ID controls (11.300). Unique ID and password combinations, periodic checks or password aging, loss management for tokens and cards, and safeguards that detect and report attempted misuse.7

The Certification Letter to FDA

One requirement is easy to miss because it is handled outside the system. Under 21 CFR 11.100(c), people using electronic signatures must, before or at the time of use, certify to FDA that the electronic signatures in their system are intended to be the legally binding equivalent of traditional handwritten signatures. The certification is signed by hand and submitted on paper or electronically, and the regulation points to FDA’s web page on Letters of Non-Repudiation Agreement for where to send it.7 That FDA page offers two sample letters, one for named individuals and one covering all of a company’s employees, and notes that a physical copy no longer needs to be mailed.15 Many companies already have a company-wide letter on file from registering for FDA’s Electronic Submissions Gateway. Ask regulatory affairs to confirm that a letter exists and what it covers before the first GMP electronic signature goes live, and file a copy with the validation package.

Right-Size the Validation Effort

Fear of a large validation project keeps many sites in print-and-sign for years. The effort does not need to be large. The electronic signature function is usually a standard, vendor-tested feature, and the site’s job is to confirm it is configured correctly and works for the intended use. FDA’s Computer Software Assurance guidance, issued February 3, 2026, is written for medical device production and quality system software, but drug manufacturers use it by analogy. It describes scaling assurance to process risk, using unscripted testing for functions that are not high risk, and relying on validation work done by developers and suppliers where appropriate.16 For a signature function, a practical package covers vendor assessment, configuration verification, testing of the specific controls above (including what happens when a signed record is changed), and the SOP and training updates.

Annex 11 sets the same expectations for EU-facing sites: electronic signatures should have the same impact as handwritten signatures within the company, be permanently linked to their record, and include the time and date they were applied. Where a computerized system records batch certification and release, Annex 11 states, “This should be performed using an electronic signature.”13

Watch for signatures that only look electronic. A scanned image of a handwritten signature pasted into a PDF, or a footnote saying “electronically signed,” is not an electronic signature. WHO states that “an inserted image of a signature or a footnote indicating that the document has been electronically signed is not adequate unless it was created as part of the validated electronic signature process.”2 MHRA says the same.4 When teams are told to stop printing, some start pasting signature images into documents instead. Name this practice in the SOP and prohibit it.

Do Not Forget the People Side

Enabling the function is half the work. The other half is making sure every person who used to sign paper can sign in the system. That means accounts for approvers who never had one, role assignments that match the SOP, training on what each signature meaning commits them to, and a clear rule that people never share credentials. A department head who has to ask a colleague to “click approve for me” will go back to paper, or worse. Plan the account and training work at the same time as the configuration, not after it.

Step Three: Put Interim Controls in Place While Paper Remains

Group C processes, and Group B processes waiting their turn, will stay hybrid for months or longer. That is acceptable only if the interim period is controlled. The Fareva Amboise warning letter is a clear example of what FDA expects. The firm had purchased new equipment with electronic data collection, planned a review of other in-process control equipment, and committed to an investment plan for “eliminating reliance on paper-based data collection systems.” FDA still found the response inadequate “because it fails to address the immediate and ongoing data integrity risks during the implementation timeline,” noting a lack of interim controls and of enhanced quality unit oversight, and FDA asked for a summary of interim controls while the CAPA plan is being carried out.12

A replacement plan is not a control. Buying a new system and setting a go-live date does not protect the data being generated today. Regulators expect interim controls that start now and last until the paper is gone. WHO puts it directly: where hybrid systems are used, including for an interim period, the risk should be assessed and mitigating controls documented.2

Write a System Description for Each Hybrid Process

PIC/S expects a detailed system description covering the whole hybrid system: its major components, the function of each, the controls for data management and integrity, and how the components interact. It also expects procedures that control the interface between manual and automated parts, including manual entry of data into systems and transcription of system data onto paper.1 For most hybrid processes this can be a few pages. Its main job is to state in plain words what the complete record is, which part is electronic, which part is paper, and how they are tied together.

Link Every Signature to Specific Electronic Files

If a handwritten signature must stay for now, make it point at the data. McDowall describes a practical approach for chromatography: a controlled blank form linked to the analysis SOP, on which the performer records the data files created, the contextual metadata used, and where the records are stored, and then signs. The reviewer uses the form as a list of electronic records to review on screen, along with the applicable audit trail entries, and no chromatograms are printed.11 The signature is then tied to a defined set of electronic records, which is much closer to what 21 CFR 11.70 requires than a signature on a stack of printouts.

Review the Electronic Data, Not Only the Paper

Change the review procedure so the second-person review covers the electronic record set: all injections or runs in the sequence, processing history, and the relevant audit trail entries. PIC/S expects procedures for reviewing hybrid data that explain how electronic and paper data are correlated to form a complete record and what approval is expected for each system’s outputs.1 The Shimoga Chemicals letter shows the result when no one reviews the electronic data: a failing result in the system and a passing result in the batch record.9

Control and Reconcile Printouts

Where printing must continue, treat printouts as controlled documents. MHRA lists “reconciliation of controlled print-outs” among the organizational measures for data governance.4 In practice that means numbered or system-stamped printouts, a rule that misprints and reprints are kept and annotated rather than thrown away, and a periodic check that the number of printouts matches the number of runs. The Granules India findings show what happens without these controls.8

Increase Quality Unit Oversight for the Interim Period

Both MHRA and PIC/S expect increased data review for hybrid systems.41 For the interim period, that can mean periodic quality unit spot checks comparing paper records with the electronic data, focused audit trail reviews for the highest-risk systems, and tracking of hybrid-related deviations as a quality metric. These checks should be written into the CAPA and should end only when the process moves out of hybrid operation.

An interim control set that holds up to inspection. For each hybrid process that will remain for more than a few weeks: a documented record of reliance, a short system description, a signature form that names the electronic files, a review procedure that covers electronic data and audit trails, controlled and reconciled printouts, increased quality unit spot checks, and a dated replacement or configuration plan under CAPA.

Step Four: Retire the Paper and Prove It Stayed Retired

The final step is the one most programs skip. A system can be configured for electronic signatures and validated, and six months later people are still printing reports “for the file.” Retiring the paper means changing the procedures, removing the printing paths, deciding what happens to the paper that already exists, and checking that the change held.

Plan the Cutover Process by Process

Cut over one process at a time, not the whole site at once. For each process, set a date after which the electronic signature is the only signature, and withdraw the paper version of the SOP, template, or worksheet on that same date. Remove the signature boxes from batch record templates for steps the system now signs. Where the system can do it, turn off or restrict the print function for reports that should no longer be printed. A printer that is still available will be used.

Decide What Happens to Legacy Paper

The paper that already exists does not disappear on cutover day. For each process, decide whether the legacy paper stays as part of the historical record, or whether it is replaced by true copies. FDA’s guidance confirms that electronic copies can serve as true copies of paper or electronic records if they preserve the content and meaning of the original, including all metadata needed to reconstruct the activity.3 MHRA allows paper data to be retained using a validated scanning process, provided there is a documented process to ensure the result is a true copy, and it asks organizations to consider any risk associated with destroying original records.4 Whatever you decide, the links between legacy paper and electronic data must be kept for the full retention period.4 Do not destroy paper originals until the true copy process is validated and the retention decision is documented.

Measure Whether the Habit Came Back

Set an effectiveness check for each cutover, as you would for any CAPA. Useful measures are simple: printouts per month from the affected system, the number of paper records arriving at the archive for that record type, and audit or walk-through findings of printed and signed copies. A follow-up walk of the same printers you checked during the inventory, three and six months after cutover, is often the most telling test.

A Sequenced Plan

Putting the four steps together gives a sequence most sites can follow over a year or two, depending on the number of systems.

1

Map and Decide (First 6 to 8 Weeks)

Build the hybrid process inventory from SOPs, templates, system lists, printers, and the archive. Sort each process into Groups A to D. Record the record of reliance for each and write it into procedures. Open a CAPA covering the full program so the plan and dates are under formal control.

2

Stop the Habit-Only Printing (Group D)

Revise SOPs for processes where the system already signs electronically. Withdraw paper templates, turn off unnecessary printing, and train staff. These changes need no system work and often remove a large share of the paper quickly.

3

Control What Remains (Groups B and C)

Put interim controls in place for every hybrid process that will continue: system description, signature-to-file linking form, electronic data review, printout reconciliation, and increased quality unit checks. Do this before any configuration work starts.

4

Enable Electronic Signatures (Group B)

Configure and validate the signature function system by system, in order of data criticality. Confirm the Part 11 controls, check that the certification letter is on file, set up accounts and roles, and train signers. Cut over each process on a fixed date.

5

Replace What Cannot Sign (Group C)

Work replacement or upgrade of systems with no signature function into capital and project plans with dates, as WHO expects. Keep interim controls running until each system is replaced.

6

Retire, Verify, and Close

Decide the fate of legacy paper, validate any scanning or true copy process, and keep paper-to-electronic links. Run effectiveness checks at three and six months. Close the CAPA only when the measures show the paper has stayed gone.

Conclusion

Print-and-sign feels safe because a handwritten signature on paper is familiar and visible. But for any system that stores or processes data, the printout is not the complete record, the signature is not linked to the data, and the review covers only what the paper shows. FDA, MHRA, PIC/S, and WHO all expect sites to define the complete record and review it, and PIC/S and WHO have set a clear direction: replace hybrid systems, with interim controls and a dated plan in the meantime. Most of the work is not technical. It is an honest inventory, a written decision about which record you rely on, procedures that match the system, and the discipline to remove the printing path once electronic signatures are in place.

Sakara Digital works with pharma and biotech organizations on data integrity, computerized system compliance, and the move from paper-based quality processes to electronic ones. If you are trying to find where your own hybrid records are, or you want an independent view on how to sequence the change, we are happy to have that conversation.